Inductively Coupled Plasma Mass Spectrometer (ICPMS)


Inductively Coupled Plasma Mass Spectrometer (ICPMS) is an import substitute and a versatile analytical instrument capable of trace / ultra-trace / isotope ratio analysis of elements in aqueous samples. It is a compact, reliable and highly sensitive analytical tool with 20x106 counts per second per part per million part of Indium and detection limit is parts per billion to parts per trillion for most of the elements. ICPMS finds numerous applications in various fields like Nuclear, Environment, Geological, Forensic; industries like Semi-conductor, Chemical/ Pharmaceutical, Petro-chemical and Research & Development institutes.

Inductively Coupled Plasma Mass Spectrometer (ICPMS), an import substitute, is a very versatile analytical instrument capable of trace and ultra-trace analysis and isotope ratio analysis in aqueous samples.

SALIENT FEATURES OF THE ICPMS

  • ICPMS is a compact, reliable and high sensitivity analytical tool consisting of an inductively coupled plasma ion source, quadrupole mass analyzer, rotary & turbomolecular pump for clean vacuum, easy sample introduction, channel electron multiplier detector and PC based interface & control.
  • Its sensitivity is 20x106 counts per second per 1 ppm of Indium and detection limits are in the range of ppt to ppb for most of the elements across the periodic table.

APPLICATIONS

  • ICPMS finds numerous applications in various fields like Nuclear, Environment, Geological, Forensic; industries like Semi-conductor, Chemical/ Pharmaceutical, Petro-chemical and Research & Development institutes.

INTRODUCTION

Inductively Coupled Plasma Mass Spectrometer (ICPMS) is a very versatile analytical instrument capable of trace and ultra-trace analysis and isotope ratio analysis in aqueous samples. It consists of an inductively coupled plasma ion source, quadrupole mass analyzer, rotary & turbomolecular pump for clean vacuum, easy sample introduction, PC based interface & control & Channel electron multiplier detector. Its sensitivity is 20x106 counts per second per 1 ppm of Indium and detection limits are ppt to ppb for most of the elements.

APPLICATIONS

ICPMS finds numerous applications in Nuclear field, Environmental fields, Geological laboratories, Semi-conductor industries, Chemical/ Pharmaceutical industries, Petro-chemical industries, Forensic laboratories and Research & Development institutes.

SPECIFICATIONS

Ion source

 
  • Source
  • :Inductively coupled plasma
     
  • RF Power
  • :Upto 2kW @ 27.12MHz
     
  • Sample introduction system
  • :Aqueous samples through nebulizer & spray chamber
     
  • Ion Optics
  • :Double deflection type with Einzel lens


    Quadrupole Mass Analyzer

     
  • Rod diameter
  • :16 mm
     
  • Rod length
  • :250 mm
     
  • Pre-filter/post filter
  • :30 mm
     
  • RF frequency
  • :1.2 MHz
     
  • Resolution
  • :1 amu (adjustable)
     
  • Mass range
  • :1-300 amu
     
  • Sensitivity
  • :10 x 106 counts /sec for 1 parts per million of Indium

    Vacuum System

     
  • Vacuum system
  • :Two turbo molecular pumps (600 l/s & 300l/s)
     
  • Roughing pump
  • :Two rotary pumps (35 m3/h & 10 m3/h)

    INFRASTRUCTURE REQUIRED

    Mechanical Fabrication Workshop with the following facilities

    • Conventional Lathe Machine
    • CNC Lathe Machine
    • Centreless Grinding Machine
    • Vertical Milling Machine
    • TIG Welding
    • Electropolishing
    • CNC jig Boring

    Facilities for Inspection of mechanical sub - assemblies

    • Normal Precision Measuring Instruments
    • 3D-Co-ordinate Measuring Machine (L.C.—1 micron)
    • MSLD ( Helium Leak Detector ) for Vacuum Testing (Leak Rate: 10-12 Torr Litre/sec)

    Electronics Assembly & Test Facilities

    • RF load, RF power meter
    • PCB assembly, SMD workbench & Instrument rack wiring facility
    • Normal measuring instruments like multimeters & 100MHz Oscilloscopes
    • Function generator ( 1MHz )
    • 6 & 1/2 Digit Digital Voltmeter or Digital Multimeter
    • R.F. Digital Voltmeter with attenuator
    • High Voltage measurement system

    Space

    • 300 square feet space is required for assembly, testing & installation of the Inductively coupled plasma mass spectrometer. Venting arrangement is required for Ozone produced by the from plasma

    Power Supply

    • 230 Volts, 32 Ampere, Three Phase, 4kVA with RF ground

    MATERIAL AVAILABILITY

    Most of the components are available within the country and some components need to be imported through local agents.

    MANPOWER

    A physics post graduate, electronics engineer, mechanical engineer & 2 trained technicians will be required for the development of this instrument.

    SR NO. Organization name Licensee Name Licensee Code ACQUIRED ON Valid Till Address City State Email Phone
    01 M/s. ELICO Ltd K.V.S.N. Raju 115 23-09-2013 22-09-2025 M/s. ELICO Ltd., B-90, APIE, Sanathnagar Hyderabad Telangana rajuk@elico.co 04023771262
    SR NO. COUNTRY NAME PATENT NUMBER
    No Data Found.
    SR NO. NAME STANDARD NUMBER LOGO
    No Data Found.

    General License Fee : ₹ 960,000.00 (Nine Lakh Sixty Thousand)

    General Royalty : 0%