EGSS is a standalone, solar powered, battery operated unit with GSM based online wireless data communication for installation in open field. It has a 2"×2" NaI(Tl) based gamma spectrometer, together with GM tube based gross gamma detectors, the former being automatically switched on when the dose rate detected by the latter increases beyond a pre-set threshold value. The establishment of a network of such field-installed systems with online communication would facilitate the real-time identification of radionuclide releases in case of radiological/nuclear emergencies. This would help in turn to facilitate prompt radionuclide-specific countermeasures in case of such incidents.
EGSS is a standalone, solar powered, battery operated unit with GSM based online wireless data communication for installation in open field. It has a 2"×2" NaI(Tl) based gamma spectrometer, together with GM tube based gross gamma detectors, the former being automatically switched on when the dose rate detected by the latter increases beyond a pre-set threshold value. The establishment of a network of such field-installed systems with online communication would facilitate the real-time identification of radionuclide releases in case of radiological/nuclear emergencies. This would help in turn to facilitate prompt radionuclide-specific countermeasures in case of such incidents.
Sr. No. | Specification | Value |
---|---|---|
1 | Size | Height: 92 cm Base area: 35 × 35 cm2 weight: 28 kg |
2 | Detectors | Gross gamma: Energy compensated GM tubes (2 redundant GM tubes for low dose rate measurement and 1 for high dose rate measurement) Spectroscopy: 2" x 2" NaI(Tl) coupled to 2" dia. bialkali PMT |
3 | Electronics | FPGA based MCA, automatic spectrometer triggering circuit based on feedback from GM tubes |
4 | Power supply | Solar Panel: 18 V/75 W Battery: 12 V/ 42 A-h lead-acid SMF battery Protection: Charge controller to protect from overcharging and deep discharge Back up: 3 days (continuous Spectroscopy), 15 days (normal operation) |
5 | Communication | SMS (GSM) |
6 | Calibration | Performed at RSS, RSSD, BARC (National Standard Laboratory) |
7 | Range | Dose rate: 50 nGy/h – 10 Gy/h (GM) Energy: up to 2 MeV |
8 | Environmental endurance | Temperature: 0 – 60 °C Relative humidity: up to 100 % Enclosure: IP68 compliant top chamber housing detectors and electronics, IP66 compliant bottom chamber housing battery (with support for solar panel) |
- Online environmental gamma Spectroscopy around
- Research reactors
- Nuclear power plants
- Fuel fabrication/reprocessing facilities
- Thermal power plants
- Oil rigs
- Shipping industries, etc.
- Network of field-installed systems with online communication to facilitate real-time identification of radionuclide releases in case of radiological/nuclear emergencies, thereby facilitating prompt radionuclide-specific countermeasures in case of such incidents.
EGSS is a standalone, solar powered, battery operated unit with GSM based online wireless data communication for installation in open field. It has a 2"×2" NaI(Tl) based gamma spectrometer, together with GM tube based gross gamma detectors, the former being automatically switched on when the dose rate detected by the latter increases beyond a pre-set threshold value. The establishment of a network of such field-installed systems with online communication would facilitate the real-time identification of radionuclide releases in case of radiological/nuclear emergencies. This would help in turn to facilitate prompt radionuclide-specific countermeasures in case of such incidents.
- Standalone, solar-powered and battery operated field-deployable system for real-time gamma Spectroscopy in open environment.
- Online wireless GSM based data communication from field-installed system to central receiving station.
- NaI(Tl)-PMT based spectrometer.
- Utilizes an in-house designed and developed FPGA based MCA.
- Only a flat surface required for installation.
- No requirement of AC mains supply.
- No requirement of on-site data logging computer.
- GM counter based ERM turns-on spectrometer only when ambient gamma dose rate goes beyond a pre-set threshold level.
- Once the battery is fully charged, works up to nearly 3 days in absence of sunlight with Spectroscopy turned-on (works for nearly 15 days in normal background radiation environment).
Sr. No. | Specification | Value |
---|---|---|
1 | Size | Height: 92 cm Base area: 35 × 35 cm2 weight: 28 kg |
2 | Detectors | Gross gamma: Energy compensated GM tubes (2 redundant GM tubes for low dose rate measurement and 1 for high dose rate measurement) Spectroscopy: 2" x 2" NaI(Tl) coupled to 2" dia. bialkali PMT |
3 | Electronics | FPGA based MCA, automatic spectrometer triggering circuit based on feedback from GM tubes |
4 | Power supply | Solar Panel: 18 V/75 W Battery: 12 V/ 42 A-h lead-acid SMF battery Protection: Charge controller to protect from overcharging and deep discharge Back up: 3 days (continuous Spectroscopy), 15 days (normal operation) |
5 | Communication | SMS (GSM) |
6 | Calibration | Performed at RSS, RSSD, BARC (National Standard Laboratory) |
7 | Range | Dose rate: 50 nGy/h – 10 Gy/h (GM) Energy: up to 2 MeV |
8 | Environmental endurance | Temperature: 0 – 60 °C Relative humidity: up to 100 % Enclosure: IP68 compliant top chamber housing detectors and electronics, IP66 compliant bottom chamber housing battery (with support for solar panel) |
- Online environmental gamma Spectroscopy around
- Research reactors
- Nuclear power plants
- Fuel fabrication/reprocessing facilities
- Thermal power plants
- Oil rigs
- Shipping industries, etc.
- Network of field-installed systems with online communication to facilitate real-time identification of radionuclide releases in case of radiological/nuclear emergencies, thereby facilitating prompt radionuclide-specific countermeasures in case of such incidents.
Space
- No requirement of large working space
- 250 sq. ft total area may be adequate
Manpower
- Two Electronics or Instrumentation Engineers for production,assembly & testing of electronic components, development of firmware and application software, etc.
- One Diploma holder in Mechanical Engineering for manufacturing of mechanical assembly
Power Supply
- For testing, a 24 V/5 Avariable power supply can be used
Test Equipment
- Standard Test & measuring facility for electronic systems, mainly oscilloscope, digital multimeter, soldering station, etc.
Other Facilities
- Microcontroller programming and testing
- FPGA program development & debugging tools and Xilinx 7 series FPGA programming facility
SR NO. | Organization name | Licensee Name | Licensee Code | ACQUIRED ON | Valid Till | Address | City | State | Phone | |
---|---|---|---|---|---|---|---|---|---|---|
01 | M/s Nuclenix Systems Pvt. Ltd. | Sh J.N Reddy | 194 | 23-06-2022 | - | - | - | - | info@nucleonix.com | |
02 | Pla Electro Appliances Pvt. Ltd. | Shri. Tarak N. Shah | 483 | 15-07-2022 | 14-07-2027 | Pla House, Thakor Estate, Kurla Kirol Road, Vidyavihar (West), Mumbai - 400086 | Mumbai | Maharashtra | plaelectro65@gmail.com | 022-25116864 |
03 | Ace-Ex Industries | Mr. Venugopal M. R. | 499 | 23-06-2022 | 22-06-2027 | Ace-Ex Industries, 248, ORM, A Wing, Royal Palms, Aarey Colony, Goregaon (E), Mumbai - 400065 | Mumbai | Maharashtra | acexindustries@gmail.com | 9222282333 |
SR NO. | COUNTRY NAME | PATENT NUMBER | |||||||||||
---|---|---|---|---|---|---|---|---|---|---|---|---|---|
No Data Found. |
SR NO. | NAME | STANDARD NUMBER | LOGO | ||||||||||
---|---|---|---|---|---|---|---|---|---|---|---|---|---|
No Data Found. |
General License Fee : ₹ 200,000.00 (Two Lakh)
General Royalty : 0%